10 2.2 - Main Menu When you power the MC10X up, it will automatically boot and start the MC10X program on the controller. The unit will also home
11 Figure 2.3.1 Volume – Any integer value from 1 to 9999uL may be selected. Speed – This adjusts the percentage of the pumps maximum speed. More
12 Orientation– Select between the Portrait or Landscape plate orientation. This is determined by which manifold you are working with.12 cha
13 Figure 2. 4.1 Volume – This is the total volume that will be primed through the system. Enter a value between 120 and 9999uL. Speed – This adj
14 Figure 2.5 1 2.6 - Auto-Prime Priming can be programmed to occur at a user specified interval. This allows the MC10X to dispense a certain a
15 Figure 2.6.1 Time – This is the interval in seconds; a time of 0 deactivates this feature Volume – This is the total volume that will be moved
16 Figure 2.7.1 Figure 2.7.2
17 Adjust Prime Position – This allows the user to jog the dispensing manifold to the position where they would like all priming functions to occur.
18 Figure 2.7.4 Figure 2.7.5 Adjust Track/Nest Height – This value should be evaluated and adjusted if manifold types or plate nests are changed.
19 plate nest without touching it. To adjust, place a sheet of paper or business card on the plate nest. Jog the manifold down until it touches the p
2 Index Introduction ... 3 Section 1 - S
20 Figure 2.7.7 Adjust Row/Column Spacing - This setting is set by the manufacturer and should not be adjusted. # Of Tips in Manifold - This setti
21 Figure 2.8.1
22 Section 3 - Maintenance of the Mirco10X Fluid Path Note: Please turn power off and disconnect all cables from the rear of the Micro10x before
23 Figure 3.1 2 Figure 3.1.3 Pivot Bearing
24 3.2 - Disassembly of the Pump Head Module After the pump head module is removed from the pump you will able to access the Teflon adapters. The
25 Figure 3.2 3 Figure 3.2.4 With the Seal nut loose you will carefully slide the piston, seals and seal nut out of the cylinder case (Figure 3.2
26 Figure 3.2.5 3.3 – Cleaning of the Pump Assembly All of the pump parts may be cleaned in an ultrasonic cleaner with a dilute detergent solutio
27 Item # QTY Part Number Description 1 1 PMP-1018 Cylinder Nut 2 1 PMP-1021 Gland Washer, ¼”. Teflon 3 2 PMP-1019 Lip Seal, ¼”, Rulon RA 4 1 PMP-101
28 Section 4 – Micro10x Validation Section 4.1 – General Validation After Cleaning To Ensure that the unit is dispensing properly after cleaning yo
29 Section 5 - Hardware Specifications Section 5.1 - General NOTE: Specifications are subject to change without notice. Micro10X: Plate
3 Introduction The Hudson Robotic Micro10x™ is a fast, compact, accurate benchtop instrument for reliable, automated bulk reagent dispensing into al
30 Liner Bore Size: ¼” Tubing Material: Peek fittings on Telfon tubing Max Shot Size 225ul per revolution Min Shot Size 12ul per rev
4 Section 1 - Setup 1.1 - Tubing The MC10X is shipped with two tube assemblies (Figure 1.1.1). The tube assembly with the two Flangeless fittings s
5 Figure 1.1.2 Figure 1.1.3 Right pump port Manifold port Left pump port
6 The priming trough uses a short silicon drain tube and waste basin. The silicon tube can be replaced with a longer one if needed. Standard silico
7 1.3 - Manifold Install the manifold with the inlet port closest to the pump head. Use the provided thumbscrews to secure the manifold ont
8 Figure 1.4.1 Controller Cable DC Inlet for Power cord Pump Cable Pump Cable Port for optional 4-way valve Serial Port Power Switch
9 Section 2 - Operation of Micro10X through the Touch Screen Controller 2.1 - Controller Features The Micro10X comes with a windows based touch sc