Hudson-robotics Micro10x Manuel d'utilisateur Page 11

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 30
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 10
11
Figure 2.3.1
Volume Any integer value from 1 to 9999uL may be selected.
Speed This adjusts the percentage of the pumps maximum speed. More viscous
liquids will require a slower pump speed to avoid cavitation.
Depth Negative values are the distance below the travel height where the tips
will move during the dispensing, allowing the user to dispense with the tips in the
wells. Positive values will move the dispensing location above the travel height.
Entering 0 will keep the travel height and the dispense height the same.
Height This is the height the tips will travel above the nest during the
dispensing. This value should be at least 1mm above the measured plate height.
The Standard Microtiter plate is ~15mm.
Wells Select from 96, 384 or 1536
Vue de la page 10
1 2 ... 6 7 8 9 10 11 12 13 14 15 16 ... 29 30

Commentaires sur ces manuels

Pas de commentaire