Figure 2.6.1
Time This is the interval in seconds; a time of 0 deactivates this feature
Volume This is the total volume that will be moved through the manifold. Enter
a value between 120 and 9999uL.
Speed This adjusts the percentage of the pumps maximum speed. More viscous
liquids will require a slower pump speed to avoid cavitation. Do not exceed more
than 30% speed on any value over 1000ul.